FDA recall Z-0427-2026

Tandem Diabetes Care, Inc. · Class I · device

Product

Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient). Pump, Tandem Mobi, Control-IQ/1010750, Pump, Tandem Mobi, Control-IQ, Replacement/1012719, Pump, Tandem Mobi, Control-IQ 7.6.0.3, Medicare/1013501, Pump, Tandem Mobi, Control-IQ 7.7/1013655, Pump, Tandem Mobi, Control-IQ, 7.7, Replacement/1013656, Pump, Tandem Mobi, Control-IQ 7.7, Medicare/1013700, Pharmacy Kit, Starter Pack, Tandem Mobi 7.7/1014081

Reason for recall

Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.

Distribution

US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, PT, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY

Key facts

Status
Ongoing
Initiation date
2025-10-06
Report date
2025-11-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0427-2026