FDA recall Z-0446-2021

Shanghai United Imaging Healthcare Co., Ltd. · Class II · device

Product

Computed Tomography X-Ray System, Model uCT 550, Rx, CE, GTIN: 06971576831036 - Product Usage: ntended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

Reason for recall

A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.

Distribution

US Nationwide distribution including in the state of TX.

Key facts