FDA recall Z-0446-2024

Elekta, Inc. · Class II · device

Product

ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation Therapy System Accelerator, Linear

Reason for recall

Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.

Distribution

worldwide

Key facts

Status
Ongoing
Initiation date
2023-10-25
Report date
2023-12-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Atlanta, GA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0446-2024