FDA recall Z-0446-2026

Medline Industries, LP · Class I · device

Product

DYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSCOPY-LF

Reason for recall

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

Distribution

International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;

Key facts

Status
Ongoing
Initiation date
2025-10-10
Report date
2025-11-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0446-2026