FDA recall Z-0448-2021

Shanghai United Imaging Healthcare Co., Ltd. · Class II · device

Product

Positron Emission Tomography and Computed Tomography System, Model uMI 550, Rx, CE, GTIN: 06971576832026 - Product Usage: intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Reason for recall

A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.

Distribution

US Nationwide distribution including in the state of TX.

Key facts

Status
Ongoing
Initiation date
2020-08-05
Report date
2020-11-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Shanghai, China

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0448-2021