FDA recall Z-0448-2022

Flower Orthopedics Corporation · Class II · device

Product

Flex-Thread Inserter Shaft intended for Fibula intramedullary-intended to be used with the Inserter Draw Rod and Blue Ratchet Handle in order to facilitate the insertion of the Flex-Thread Nail System- Catalog Number: 8435-1

Reason for recall

Mating incompatibility between the Inserter Shaft and the Inserter Draw Rod causing the inserter draw rod to bind and not rotate freely as intended surgeon may be unable to complete surgery unless a second set is available

Distribution

US Nationwide distribution in the states of AZ, CA, CO, GA, TN, FL, IL, MA, MD, NC, TX.

Key facts

Status
Terminated
Initiation date
2021-11-02
Report date
2022-01-12
Termination date
2023-10-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Horsham, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0448-2022