FDA recall Z-0452-2022

Biomeme, Inc. · Class II · device

Product

Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555; SARS-CoV-2 Go-Plates, 3000562; SARS-Cov-2 Bulk Vials, 3000564

Reason for recall

The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the RPC Buffer once resuspended from the current instruction of one week to two days when stored at room temperature (15degC-30degC).

Distribution

Domestic distribution to AZ, CA, FL, HI, IN, KY, MD, MO, NC, NJ, NY, OH, OK, PA, TX, VA. Foreign distribution to Ghana, Dominican Republic, and Singapore.

Key facts

Status
Terminated
Initiation date
2021-10-20
Report date
2022-01-12
Termination date
2024-03-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Philadelphia, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0452-2022