FDA recall Z-0453-2026

GE Healthcare Finland Oy · Class II · device

Product

Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Record updated 3/19/26 to include: CARESCAPE B450 MBA313 2095800-001 CARESCAPE MONITOR B450 (ATO MODEL) 2068491-002 CARESCAPE MONITOR B450 B1 2107633-001 CARESCAPE MONITOR B450-GS (V2.0.8) 2094080-001

Reason for recall

Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.

Distribution

Worldwide - US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2025-09-26
Report date
2025-11-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Helsinki, Finland

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0453-2026