FDA recall Z-0462-2019

Invivo Corporation · Class II · device

Product

Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388

Reason for recall

The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard

Distribution

Distributed to the Netherlands

Key facts

Status
Completed
Initiation date
2018-09-24
Report date
2018-11-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Orlando, FL, United States