FDA recall Z-0473-2023

Becton Dickinson & Co. · Class II · device

Product

BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species. Catalog (Ref) Number: 221267

Reason for recall

Boxes contain a reduced concentration of Nicotinamide Adenine Dinucleotide and does not provide enough of the blood factors necessary to support Haemophilus growth. The reduction could cause a false negative result for Haemophilus isolation in a clinical specimen.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, India, Philippines.

Key facts

Status
Ongoing
Initiation date
2022-11-21
Report date
2022-12-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Sparks, MD, United States