FDA recall Z-0477-2021

Luminex Corporation · Class II · device

Product

Verigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. Used with assay Luminex, VERIGENE Gram-Negative Blood Culture Nucleic Acid Test, BC-GN, IVD - Product Usage: IVD sample processor, which utilizes a single-use disposable test consumable and a self-contained Verigene Test Cartridge for each sample tested.

Reason for recall

Luminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in some cases, were manufactured using unvalidated processes. These text fixtures were used to calibrate and verify VERIGENE SP devices.

Distribution

Worldwide distribution. US nationwide and Kuwait, Austria, United Kingdom, Spain, France, Chile, Japan, Netherlands, Greece, Italy, Thailand, Germany, Turkey

Key facts

Status
Terminated
Initiation date
2020-10-13
Report date
2020-12-02
Termination date
2024-01-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northbrook, IL, United States

Primary source