1. Sensis, material # 10764561; UDI: 04056869010137 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205 The Sensis recording system is a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac, interventional Radiology, and surgical studies.
Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain sporadic circumstances, the CO (Cardiac Output) measurement using the Thermodilution method will temporarily no longer be possible.
US Nationwide distribution in the states of AK, AL, AR, AZ, CA, FL, GU, HI, IA, IL, IN, KS, KY, MA, MI, MO, MS, NC, NE, NJ, NY, OH, OK,PA, PR, SD, TN, TX, WA, WV.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0499-2022