FDA recall Z-0515-2021

Covidien Llc · Class II · device

Product

Capnostream 35 Portable Respiratory Monitor, Product numbers PM35MN02, PM35MN01, PM35MN05, DLPM35MN02 - Product Usage: use during both no motion and motion conditions and for patients who are well or poorly perfused.

Reason for recall

The firm has released software update V01.05.02.16 (also known as V1.5.2) in response to customer reports of a false display of the message, "Temperature Exceeds Limits" followed by automatic shutdown of the monitor with no accompanying alarm.

Distribution

Worldwide distribution - US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2020-11-16
Report date
2020-12-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mansfield, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0515-2021