FDA recall Z-0526-2018

Angiodynamics, Inc. · Class II · device

Product

Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600, Model # (UPN) H7879006000, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

Reason for recall

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Distribution

Worldwide distribution - US Nationwide

Key facts

Status
Terminated
Initiation date
2017-06-05
Report date
2018-02-14
Termination date
2018-09-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Queensbury, NY, United States

Primary source