# FDA recall Z-0527-2019

> **Barco N.V.** · Class II · device recall initiated 2017-03-30.

## Product

Optional MDRC-1119 Front Cover Attachment Kit  for review display monitors.    Product Usage:  The product is an optional kit, intended to be attached in front of the LCD panel of a review display (MDRC-1119).

## Reason for recall

If metal strips are removed and re-applied to another display, there is potential for the metal strips to be deformed resulting in a smaller contact surface with the front cover magnet strips.  There is a risk that the protective front cover could come loose and may fall down.

## Distribution

Worldwide - US Nationwide Distribution to CA, IL, GA, TX, and WI. In the countries of Belgium, Brazil, Canada, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Poland, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Taiwan, Turkey.

## Key facts

- **Recall number:** Z-0527-2019
- **Recalling firm:** Barco N.V.
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2017-03-30
- **Report date:** 2018-12-12
- **Termination date:** 2020-05-04

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Kortrijk, N/A, Belgium