FDA recall Z-0529-2022

Baxter Healthcare Corporation · Class I · device

Product

Baxter Spectrum IQ Infusion Pumps, Product code 3570009.

Reason for recall

There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.

Distribution

Worldwide Distribution. US nationwide Puerto Rico and St. Thomas. There was government/military distribution. Countries of Canada, Bahamas, Barbados, Bermuda, Guyana, Jamaica, and Trinidad and Tobago.

Key facts

Status
Ongoing
Initiation date
2021-12-29
Report date
2022-02-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Deerfield, IL, United States