FDA recall Z-0532-2021

Canon Medical System, USA, INC. · Class II · device

Product

CAS-930A Flat Panel Monitor (FPD) used in conjunction with the Canon Interventional Angiography System, Model numbers: Infinix-8000C and Infinix-8000H - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Reason for recall

There is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in a delay with procedure and/or potential damage to the patients blood vessels due to removal without visibility.

Distribution

US Nationwide distribution including in the states of LA, CA, CO, WV, GA, NH, FL, PA, WI, MN, ID.

Key facts

Status
Terminated
Initiation date
2020-10-08
Report date
2020-12-09
Termination date
2022-12-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tustin, CA, United States

Primary source