FDA recall Z-0540-2022

DIXI MEDICAL USA · Class II · device

Product

The DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrode that is 0.8 mm in diameter with a rounded (hemispherical) tip. The electrode has multiple platinum/iridium contacts that are 2 mm long, with a total exploration length varying from 16 mm to 80.5 mm according to the electrode reference. Microdeep Depth Electrodes are intended for temporary use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain.

Reason for recall

Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag

Distribution

US Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2021-12-09
Report date
2022-02-09
Termination date
2023-09-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Plymouth, MI, United States

Primary source