FDA recall Z-0545-2022

Immuno-Mycologics, Inc · Class II · device

Product

CRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detection of Cryptococcal Antigen in serum and CSF, REF CR2003, IVD, UDI: (01)00816387020919

Reason for recall

The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive test results may be false positives, which may cause some patients to initiate unnecessary anti-fungal therapy. Samples with negative test results are NOT affected.

Distribution

US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA.

Key facts

Status
Terminated
Initiation date
2022-01-03
Report date
2022-02-09
Termination date
2024-05-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Norman, OK, United States