FDA recall Z-0548-2018

Zimmer Biomet, Inc. · Class II · device

Product

M/DN Intramedullary Fixation Humeral Guide Wire - Bullet Tip 2.4 mm Diameter 70 cm Length Single Use Only, Item Number 00-2255-026-00 Product Usage: The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wire incorporates either a ball or tapered tip to aid in removal of the intramedullary reamer if it becomes lodged or fails.

Reason for recall

The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.

Distribution

Nationwide within U.S.A and Puerto Rico; International: Australia, Brazil, Canada, China, Germany, India, Japan, Netherlands & Singapore

Key facts

Status
Terminated
Initiation date
2017-06-05
Report date
2018-02-14
Termination date
2019-04-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Warsaw, IN, United States