FDA recall Z-0548-2024

Integra LifeSciences Corp. · Class I · device

Product

CereLink ICP Extension Cable (Extension Cable), Model No. 826845 Used with Codman CereLink Intracranial Pressure (ICP) Monitor, Model No. 826820/826820P. Cable is also provided with monitor as part of the same SKU.

Reason for recall

Firm is expanding the existing recall RES 90457 for the Cerelink, initiated in June 2022, to include all CereLink ICP Extension Cables. The updated corrective action includes a design change to the extension cable to address the "out of range readings" issue.

Distribution

US distribution nationwide to the following states: AL AR AZ CA CO CT DC DE FL GA IA KY LA MA ME MN ND NE NH NJ NY OK PA RI TN TX UT VA WI WV OUS distribution to Australia/New Zealand, Canada, and Europe/Middle East/Africa.

Key facts

Status
Ongoing
Initiation date
2023-10-05
Report date
2023-12-20
Voluntary/Mandated
Voluntary: Firm initiated
Location
Princeton, NJ, United States