FDA recall Z-0552-2019

Medtronic Navigation, Inc. · Class I · device

Product

StealthStation Cranial and Synergy Cranial Depth Gauge: StealthStation Cranial Optical Pocket Guide - Instructions for cranial procedures using optical localization StealthStation Cranial version 3.0, 3.1 CE 0344 Rx Only 2018-02 9735489 Revision 3 Medtronic Navigation, Inc. Biopsy Needle Kit (9733068) Note: This device is an accessory to your StealthStation System. . . The Biopsy Needle is a presterilized , single-use, side-cutting needle for cranial tissue biopsies. CE 0344 Rx Only 2017-05 9733068-01 Medtronic Navigation, Inc.

Reason for recall

An incorrect display of the Biopsy Depth Gauge graphical representation in the Guidance View using Cranial software, which resulted in the biopsy of healthy tissue.

Distribution

US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY OUS: Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bosnia, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, French Guiana, Germany, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Moldova, Morocco, Myanmar, Nepal, Netherland, New Zealand, Nicaragua, Norway, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerl

Key facts

Status
Terminated
Initiation date
2018-09-26
Report date
2019-01-09
Termination date
2024-09-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Louisville, CO, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0552-2019