FDA recall Z-0553-2022

Luminex Corporation · Class II · device

Product

ARIES HSV 1&2 Assay, Catalog/REF# 50-10017

Reason for recall

Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.

Distribution

Worldwide distribution - US Nationwide distribution in the states of New Jersey, California, Connecticut, Minnesota, Missouri, New Jersey, New Mexico, Pennsylvania, Tennessee, Washington, and Wisconsin) and the country of Malaysia.

Key facts

Status
Terminated
Initiation date
2021-11-17
Report date
2022-02-09
Termination date
2024-07-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Austin, TX, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0553-2022