# FDA recall Z-0553-2025

> **Shimadzu Corp. Analytical & Measuring Instruments Div.** · Class II · device recall initiated 2024-10-09.

## Product

Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL

## Reason for recall

There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.

## Distribution

Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries of DE, CN, JP, UY.

## Key facts

- **Recall number:** Z-0553-2025
- **Recalling firm:** Shimadzu Corp. Analytical & Measuring Instruments Div.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-10-09
- **Report date:** 2024-12-04

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Kyoto, Japan

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0553-2025

## Citation

> AI Analytics. FDA recall Z-0553-2025. Retrieved 2026-07-26 from https://api.ai-analytics.org/recall/Z-0553-2025. Source: US FDA. Licensed CC0.

---