# FDA recall Z-0563-2021

> **Spacelabs Healthcare, Ltd.** · Class II · device recall initiated 2020-11-06.

## Product

Sentinel V10.x & V11.x, Models: 98200 & 98201 - Product Usage: designed using the latest programming technology and is easy to install and to use.

## Reason for recall

During system upgrade incorrect comments may be added to existing tests. In such instance: Resting ECG and ABP tests which did not contain comments at the time of the upgrade, may have unrelated comments added. Existing comments will not be overwritten or removed. Rhythm ECG Tests may have comments overwritten with unrelated comments. Any tests created after the upgrade are not affected.

## Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of PR, MN, NE, FL, IL, CA, NY, ND, AR, WI, SD, TN, MO, MA, TX, MT, OH, MI, IA, NJ, CT, NC, UT, WV, GA, CO, PA, AL, DC, NV, DE, WA, OR, IN, SC, HI, MS, KS, ID, MD, LA, NH, VA, ME, AK, VT, OK, AZ, KY, NM, RI and the countries of    AFGHANISTAN, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BELGIUM, BOLIVIA, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, ICELAND, INDIA, IRAN, IRELAND, ISRAEL, ITALY, KENYA, KOREA, REPUBLIC OF, KUWAIT, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, OMAN, PAKISTAN, PALESTINE, PANAMA, PARAGUAY, PERU, POLAND, PORTUGAL, QATAR, Republic of Macedonia, SAUDI ARABIA, Serbia, SINGAPORE, Slovakia, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED ARA

## Key facts

- **Recall number:** Z-0563-2021
- **Recalling firm:** Spacelabs Healthcare, Ltd.
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2020-11-06
- **Report date:** 2020-12-16
- **Termination date:** 2024-10-18

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Hertford, N/A, United Kingdom

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0563-2021