FDA recall Z-0564-2018

Kalila Medical · Class II · device

Product

Vado Bi-Directional Steerable Sheath 8.8F, REF SS8FMB74, packaged in a lidded tray that is within a pouch which is inserted into a shelf carton, Sterile, Rx. The firm name on the label is Kalila Medical, Campbell, CA.

Reason for recall

Torn sheath liner material was detected at the proximal end of the device during internal product testing which may be the result of damage incurred during the manufacturing process.

Distribution

Distribution was made to DC, FL, GA, MD, MI, NJ, NY, OH, PA, SC, and VA. There was also military/government distribution. There was no foreign distribution.

Key facts

Status
Terminated
Initiation date
2017-12-14
Report date
2018-02-21
Termination date
2018-12-14
Voluntary/Mandated
Voluntary: Firm initiated
Location
Campbell, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0564-2018