FDA recall Z-0569-2018

B. Braun Medical, Inc. · Class II · device

Product

Pencan Spinal Needle Tray. Product Catalog Number: 333865. Local anesthesia of the skin prior to insertion of the regional anesthesia needle.

Reason for recall

B. Braun Medical Inc. (BBMI) initiated this recall of various BBMI convenience kits containing the 22GA 1-1/2in SafetyGlide Needle due to the presence of loose polypropylene foreign matter was above release specification. This foreign matter has been identified as a product hub material which has been tested for biocompatibility per ISO 10993 during the product development process.

Distribution

US Nationwide Distribution.

Key facts

Status
Terminated
Initiation date
2017-06-21
Report date
2018-02-21
Termination date
2018-04-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Allentown, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0569-2018