FDA recall Z-0569-2021

Ra Medical Systems, Inc. · Class II · device

Product

DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in occlusive peripheral vascular disease. Indications: For crossing chronic total occlusions in patients with symptomatic infrainguinal lower extremity vascular disease.

Reason for recall

Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.

Distribution

US Distribution to states of: AL, AZ, CA, FL, IL, IN, KY, LA, MD, NC, NV, NY, OK, OR, TX, and WA; and to OUS country: Japan.

Key facts

Status
Terminated
Initiation date
2020-10-16
Report date
2020-12-23
Termination date
2023-01-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Carlsbad, CA, United States