FDA recall Z-0580-2025

Micro-X Ltd. · Class II · device

Product

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Reason for recall

Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.

Distribution

US: TX, PR, MD, CA

Key facts

Status
Ongoing
Initiation date
2024-10-22
Report date
2024-12-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tonsley, Australia

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0580-2025