FDA recall Z-0583-2020

GN Hearing A/S · Class II · device

Product

Behind-the-Ear hearing aids, Model Numbers: Q962-DRW, T561-DRW, T562-DRW, T762-DRW, T777-DW, AMZ964-DRW, ET588-DWH, ET598-DW, ET788-DWH, ET798-DW, ET988-DWH, ET998-DW, FT861-DRW, FT862-DRW, RE561-DRWC, RE762-DRW, RE961-DRWC, RE962-DRW, PZ862-DRW, LT561-DRW, LT561-DRWZ, LT562-DRW, LT588-DW, LT761-DRW, LT761-DRWZ, LT762-DRW, LT777-DW, LT788-DW, LT961-DRW, LT961-DRWZ, LT962-DRW, LT967-DW, LT977-DW, LT988-DW, TST1763-DRW, TST1764-DRW, TST663-DRW, TST963-DRW, TST964-DRW, VA861-DRW, VA861-DRWZ, VA862-DRW

Reason for recall

GN Hearing received reports from our manufacturing site in US regarding a software bug in our service & repair tool which affected the capture and restoration of user data in a limited number of repaired hearing aids. The software bug caused these repaired hearing aids to contain altered specifications for sound amplification, potentially putting users residual hearing at risk of experiencing unintended additional available gain to varying degrees i.a.18dB flat linear gain or potentially in worst case Full On Gain. However, no injuries have been reported.

Distribution

Nationwide

Key facts

Status
Terminated
Initiation date
2019-09-18
Report date
2019-12-04
Termination date
2023-05-31
Voluntary/Mandated
Voluntary: Firm initiated
Location
Ballerup, Denmark

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0583-2020