The LeadCare¿ II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole blood. The LeadCare¿ II Test Kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 packages of 24 tubes each; 48 tests) " Heparinized Capillary Tubes/Plungers (50) " Transfer Droppers (50) " Calibration Button (1) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.
US Nationwide distribution.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0584-2026