# FDA recall Z-0587-2026

> **Pivotal Health Solutions, Inc.** · Class II · device recall initiated 2025-10-09.

## Product

Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table

## Reason for recall

The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.

## Distribution

Distribution to US nationwide, Canada and Japan.

## Key facts

- **Recall number:** Z-0587-2026
- **Recalling firm:** Pivotal Health Solutions, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-10-09
- **Report date:** 2025-12-03

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Watertown, SD, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0587-2026

## Citation

> AI Analytics. FDA recall Z-0587-2026. Retrieved 2026-08-11 from https://api.ai-analytics.org/recall/Z-0587-2026. Source: US FDA. Licensed CC0.

---