FDA recall Z-0588-2019

GE Healthcare, LLC · Class II · device

Product

GE LOGIQ E10 Ultrasound System, Model #530000E10. The device is intended for use by a qualified physician for ultrasound evaluation of Fetal/Obstetrics; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculoskeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (vascular).

Reason for recall

There is a potential for an undetected over temperature condition on the trans-esophageal (TEE) probe leading to a possible burn injury.

Distribution

Worldwide Distribution: US (nationwide) to states of: AZ, IL, MA, ME, MN, MT, PA, TX, and WA and countries of: Foreign distribution was made to Austria, Denmark, Finland, and Norway.

Key facts

Status
Terminated
Initiation date
2018-09-14
Report date
2018-12-19
Termination date
2019-05-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukesha, WI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0588-2019