FDA recall Z-0592-2019

Philips Electronics North America Corporation · Class II · device

Product

IntelliVue MX40 Patient Monitor, Model Nos. 865350, 865351, 865352, 867146 - Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals.

Reason for recall

A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2018-10-10
Report date
2018-12-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Andover, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0592-2019