FDA recall Z-0595-2018

Hologic, Inc · Class II · device

Product

Hologic Panther(R) Fusion(TM), Catalog number ASY-09600 Product Usage: The Panther instrument is a fully automated sample-to-result instrument, eliminating the need for batch processing and automates all aspects of nucleic acid testing on a single, integrated platform. The Panther can be test samples with random access or batch managed; can perform multiple assays on the same patient sample with up to 4 different assays simultaneously.

Reason for recall

The Panther or the Panther Fusion glass panel separates from the 19 inch touch screen monitor and could potentially fall and hit an operator.

Distribution

Worldwide distribution

Key facts

Status
Terminated
Initiation date
2018-01-20
Report date
2018-02-21
Termination date
2022-07-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States

Primary source