FDA recall Z-0610-2020

Smiths Medical ASD Inc. · Class I · device

Product

Medfusion Syringe Pump, Model 4000. The Model 4000 pump is a small, lightweight and portable syringe infusion pump. The pump offers a variety of delivery modes programmable to meet specific patient care needs.

Reason for recall

There is an anomaly in Medfusion 4000 Syringe Pump Firmware version 1.7.0 that could potentially result in loss of primary audible and visual battery alarm functionality, and interruption of therapy.

Distribution

CO, MD, NY

Key facts

Status
Terminated
Initiation date
2019-10-28
Report date
2019-12-18
Termination date
2024-06-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Minneapolis, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0610-2020