# FDA recall Z-0613-2019

> **Tosoh Bioscience Inc** · Class II · device recall initiated 2017-03-22.

## Product

AIA-360 Analyzer is capable of performing two methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay. An antigen-antibody reaction begins by combining a patient sample, control, or calibrator with a diluent in an immunoreaction test cup from the AIA-PACK reagent series.

## Reason for recall

The firm has become aware of a potential issue with the AIA-360 analyzer that could prevent the table home sensor from detecting the table home position. If the malfunction with the table home sensor occurs on the AIA-360, the sensor board may need to be replaced. If the home sensor fails to detect the home position, the AIA-360 analyzer will cease to function and will be unavailable for use.  If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received three (3) complaints related to this issue with no serious injuries reported.

## Distribution

Nationwide distribution. International distribution to Honduras, Caribbean, Chile, Colombia, Costa Rica, Uruguay, Ecuador, Panama, Mexico, Canada, and Venezuela.

## Key facts

- **Recall number:** Z-0613-2019
- **Recalling firm:** Tosoh Bioscience Inc
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2017-03-22
- **Report date:** 2018-12-26
- **Termination date:** 2020-12-09

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Grove City, OH, United States