FDA recall Z-0613-2020

Integra LifeSciences Corp. · Class II · device

Product

C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Reason for recall

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operator s manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Distribution

US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK

Key facts

Status
Terminated
Initiation date
2019-10-21
Report date
2019-12-18
Termination date
2020-05-07
Voluntary/Mandated
Voluntary: Firm initiated
Location
Plainsboro, NJ, United States