# FDA recall Z-0621-2025

> **Becton Dickinson & Co.** · Class II · device recall initiated 2024-11-06.

## Product

BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens.     Catalog Number: 231263

## Reason for recall

BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility

## Distribution

Worldwide - US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NE, PA and the countries of Canada, Belgium, Singapore, Malaysia.

## Key facts

- **Recall number:** Z-0621-2025
- **Recalling firm:** Becton Dickinson & Co.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-11-06
- **Report date:** 2024-12-11

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Sparks, MD, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0621-2025

## Citation

> AI Analytics. FDA recall Z-0621-2025. Retrieved 2026-07-08 from https://api.ai-analytics.org/recall/Z-0621-2025. Source: US FDA. Licensed CC0.

---