FDA recall Z-0629-2018

Cincinnati Sub-Zero Products LLC · Class II · device

Product

Hemotherm¿ Dual Reservoir Cooler/Heater System Models 400CE and 400MR Product Usage: A dual reservoir, water re-circulating device that supplies temperature controlled water to a heat exchanger. The device, in conjunction with a heat exchanger, is used to warm or cool the blood that enters and exits a blood heat exchanger for adult, pediatric or infant patients.

Reason for recall

Due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust cleaning & disinfection procedure.

Distribution

US Nationwide Distribution

Key facts

Status
Terminated
Initiation date
2017-11-27
Report date
2018-02-21
Termination date
2018-07-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
Cincinnati, OH, United States

Primary source