FDA recall Z-0644-2018

Innovasis, Inc · Class II · device

Product

Excella II Polyaxial Pedicle Screw used as a pedicle screw system in skeletally mature patients, is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in treatment of the following acute and chronic deformities of the thoracic, lumbar, and sacral spine:

Reason for recall

Nine lots of 8.5 mm Polyaxial Pedicle Screws are at risk of breakage.

Distribution

United States

Key facts

Status
Terminated
Initiation date
2018-01-08
Report date
2018-02-28
Termination date
2018-02-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Salt Lake City, UT, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0644-2018