FDA recall Z-0662-2022

Respironics California, LLC · Class I · device

Product

Philips Respironics V60 Ventilator Part Number 1053617

Reason for recall

A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient. This failure may either activate both visual and audible alarms, or it may not sound or display an alarm (a silent shutdown).

Distribution

U.S.: AR, CA, FL, GA, IN, LA, MD, MO, NY, PA, SC, TN and TX. O.U.S. N/A

Key facts

Status
Ongoing
Initiation date
2022-01-24
Report date
2022-03-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Carlsbad, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0662-2022