# FDA recall Z-0671-2025

> **CareFusion 303, Inc.** · Class II · device recall initiated 2024-11-14.

## Product

BD Pyxis MedStation ES 7 Drawer Auxiliary Tower, REF: 324

## Reason for recall

Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.

## Distribution

Worldwide - US Nationwide distribution including in the states of OK, LA, FL, AR, VA, KY, PA, AL, NC, CA, MN, AK, TX, MI, ME, OH, OR, NY, IL, RI, GA, WI, NH, KS, CO, WA, MO, MD, AZ, SC, NV, WV, NM, IN, NJ, TN, DE, MT, MA, VT, DC, NE, CT, ID, WY, MS, UT, HI, IA, SD, ND, GU, MP, PR and the countries of CA, BR, JP, ES, NZ, AU, FR, AR, MX, IT, CL, TR, SA, MY, CH, IE, TH, SG, KW, BE, DE, PT, QA, MA, FI, AE, EG, BS, GB, BM, GR, BH, KR, MC, PL, LB, HK, IN, IL, MO, IQ, TN, NO, PH, CO, TW, CN, JO, OM.

## Key facts

- **Recall number:** Z-0671-2025
- **Recalling firm:** CareFusion 303, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-11-14
- **Report date:** 2024-12-18

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** San Diego, CA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0671-2025

## Citation

> AI Analytics. FDA recall Z-0671-2025. Retrieved 2026-07-15 from https://api.ai-analytics.org/recall/Z-0671-2025. Source: US FDA. Licensed CC0.

---