FDA recall Z-0676-2024

Bard Peripheral Vascular Inc · Class II · device

Product

BD Recanalization System, REF: BDRECANSYSTEM

Reason for recall

Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute warning/5-minute restart required messages to indicate the ten-minute maximum catheter use time has been exceeded, which may increase the risk of the device overheating or breaking.

Distribution

US: CA, OH, PA, NY, TX, NJ, MS, IL, GA, AL, AZ, NC, NV, PR, IN, MA, WI, KS, AR, FL, MD, IA, NE, LA, HI, MO, DE, DC, VA, NM, CO, KY, ME, MN

Key facts

Status
Ongoing
Initiation date
2023-11-14
Report date
2024-01-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tempe, AZ, United States

Primary source