FDA recall Z-0679-2025

Noah Medical · Class II · device

Product

Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)

Reason for recall

Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.

Distribution

US Nationwide Distribution: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC

Key facts

Status
Ongoing
Initiation date
2024-10-25
Report date
2024-12-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Carlos, CA, United States