FDA recall Z-0680-2025

Spectranetics Corporation · Class II · device

Product

Turbo Elite Laser Atherectomy Catheter Model Number / UDI-DI code / Product Description 414-151 / 989930000531 / 1.4 mm OTW 417-152 / 989930000551 / 1.7 mm OTW 420-006 / 989930000571 / 2.0 mm OTW 423-001 / 989930000591 / 2.3 mm OTW 425-011 / 989930000601 / 2.5 mm OTW 423-135-01 / 989930004171 / 2.3 mm OTW 423-135-02 / 989930004181 / 2.3 mm OTW 425-135-01 / 989930004201 / 2.5 mm OTW 425-135-02 / 989930004211 / 2.5 mm OTW The Turbo-Elite devices are indicated for use in the treatment, including atherectomy, of infrainguinal stenoses and occlusions in humans.

Reason for recall

Due to complaints received for the marker band detaching from the distal tip of the catheter and remaining in the patient as laser catheter was retracted.

Distribution

U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, ILIN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD TN, TX, UT, VA, WA, WI, and WV. O.U.S.: Australia, Belgium, Brazil, China, Croatia, Cyprus, Czech Republic, Denmark, France, Germany, Italy, Japan, Kuwait, Netherlands, Poland, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.

Key facts

Status
Ongoing
Initiation date
2024-11-25
Report date
2024-12-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Colorado Springs, CO, United States

Primary source