FDA recall Z-0682-2020

Polymer Technology Systems · Class II · device

Product

A1CNow+ Systems (professional use product), Software Version Revision D, REF Numbers US: 3021, 3024; EU: 3037, 3038, 3057, 3058

Reason for recall

A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.

Distribution

The products were distributed to the following US states: AR, CA, FL, IL, IN, KY, MN, NC, NJ, NY, OH, PA, RI TN, UT, VA, WA, and Puerto Rico. The products were distributed to the following foreign countries: Australia, Bahamas, Belgium, Bahrain, Bangladesh, Canada, China, Columbia, Czech Republic, England, Egypt, Greece, Spain, India, Indonesia, Haiti, Hong Kong, Italy, Japan, Jordan, Kenya, Lebanon, Mexico, Myanmar, Malaysia, Philippines, Pakistan, South Africa, Nigeria, Poland, Saudi Arabia, Sweden, Turkey, Singapore, Trinidad & Tobago, Slovenia, Vietnam, United Arab Emirates, Uruguay.

Key facts

Status
Terminated
Initiation date
2016-12-02
Report date
2019-12-25
Termination date
2020-07-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Whitestown, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0682-2020