FDA recall Z-0682-2025

Nova Biomedical Corporation · Class I · device

Product

StatStrip Glucose Ketone Hospital Meter System. Model/Catalog Number: 63683. Prescription use blood glucose ketone meter for near-patient testing

Reason for recall

A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.

Distribution

Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.

Key facts

Status
Ongoing
Initiation date
2024-11-20
Report date
2025-01-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waltham, MA, United States