FDA recall Z-0692-2022

Beckman Coulter Inc. · Class II · device

Product

AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850

Reason for recall

Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magnesium patient samples. The impact is only to the clinical interpretation of magnesium results in the presence of lipemia.

Distribution

US: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming; and OUS: Albania, Algeria, American Samoa, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Ethiopia, Finland ,France, Georgia, Germany, Ghana, Gibraltar, Greece, Guam, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Ital

Key facts

Status
Terminated
Initiation date
2021-12-08
Report date
2022-03-09
Termination date
2025-01-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Brea, CA, United States