FDA recall Z-0693-2025

Noah Medical · Class II · device

Product

Galaxy System (GAL-001)

Reason for recall

Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.

Distribution

US: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NY

Key facts

Status
Ongoing
Initiation date
2024-10-25
Report date
2024-12-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Carlos, CA, United States

Primary source